What is the difference in calculations between JP14 and JP Plate Count in Empower 2 and Empower 3? - WKB9779

Article number: 9779

ENVIRONMENT

  • Empower 3
  • Empower 2
  • System suitability
  • Japanese Pharmacopoeia

ANSWER

  1. In Empower 3, the JP 15 Plate Count field was renamed to JP Plate Count and the JP Plate Count field was renamed to JP 14 Plate Count.
  2. In Empower 3, the plate count value that is calculated with the 5.54 constant is called JP Plate Count and the plate count value that is calculated with the 5.55 constant is called JP 14 Plate Count.
  3. In Empower 2, the plate count value that is calculated with the 5.54 constant is called Plate Count and the plate count value that is calculated with the 5.55 constant is called JP 15 Plate Count.
  4. The Help text has been updated in Empower 3 FR4 to reflect these differences.

ADDITIONAL INFORMATION

We received reports from customer that Empwer 3 FR5 SR4, Empower 3.6.1 have not had the Help text updated.
CRI-4668 has been created.