SYMPTOMS

  • No data collected from CAD

ENVIRONMENT

  • Empower 3
  • Thermo CAD
  • 2695 Separations Module

CAUSE

No Inject Start or RS232 cable

FIX or WORKAROUND

Add inject start and RS232 cables.