With Empower, Thermo CAD detector does not collect data - WKB97246
Article number: 97246
SYMPTOMS
- No data collected from CAD
ENVIRONMENT
- Empower 3
- Thermo CAD
- 2695 Separations Module
CAUSE
No Inject Start or RS232 cable
FIX or WORKAROUND
Add inject start and RS232 cables.
ADDITIONAL INFORMATION
id97246, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP