Peak is missing from Empower result peak table - WKB95068
Article number: 95068
SYMPTOMS
- Peak is missing from Empower result peak table
ENVIRONMENT
- Empower
- Breeze 2
CAUSE
In the processing method, the peak retention time was set during the Inhibit integration window time, which prevented it from being identified.
FIX or WORKAROUND
Modify the processing method Inhibit integration window time.