In the Preview window for the Results of Empower 3, the Resolution field, Tailing field and the Plate Count field are blank. - WKB93180
Article number: 93180
SYMPTOMS
- When previewing the processing result, the Retention Time and Area value are shown, but the Resolution field and the Plate Count field are blank.
- It is still blank when confirming on the Review window for the Results
- System suitability option installed
- System suitability is checked in Project Properties > Enabled Options
ENVIRONMENT
- Empower 3
CAUSE
The system suitability is not enabled in the processing method.
FIX or WORKAROUND
- Review the results.
- Click Window > Processing Method.
- Click the Suitability tab.
- Select the check box of "Calculate Suitability Results". If necessary, also select the check box of "Calculate Suitability Results for Unknown Peaks".
- Enter the Void Volume Time. The Void Volume Time value is used to calculate capacity factor and selectivity.
- Save the processing method from the File menu.
- Process the data from the Sample Set (or Injection, Channel) using the saved processing method.