Result set exists when there is no processing method associated with the method set in Empower - WKB9251
Article number: 9251
SYMPTOMS
- Sample set produces a result set, but there is no associated processing method.
ENVIRONMENT
- Empower 3 Enterprise
CAUSE
The Shutdown method was launched as a sample set, with process set to "Normal".
FIX or WORKAROUND
To avoid creating an empty results set, do one of the following:
- Set the function set to "Condition" in the sample set.
- Select the Shutdown instrument method when run is launched.