Result set exists when there is no processing method associated with the method set in Empower - WKB9251

Article number: 9251

SYMPTOMS

  • Sample set produces a result set, but there is no associated processing method.

ENVIRONMENT

  • Empower 3 Enterprise

CAUSE

The Shutdown method was launched as a sample set, with process set to "Normal".

FIX or WORKAROUND

To avoid creating an empty results set, do one of the following:

  1. Set the function set to "Condition" in the sample set.
  2. Select the Shutdown instrument method when run is launched.

ADDITIONAL INFORMATION

See KCS article #2931, How to Create a shutdown method in Empower.