Report Publisher error when saving a report to PDF in Citrix - WKB91764
Article number: 91764
SYMPTOMS
- Report Publisher error when saving a report to PDF in Citrix
- Error: An Error making the file file:V:\Folder\Result1000.pdf. Please see the message center for additional details on the error.

- There is no additional information in the message center
- PDF is usually created, or it has not been shown that the file is not created
- Overwriting a previously created PDF does not give the error
ENVIRONMENT
- Empower
- Citrix
- Report Publisher
CAUSE
Cause of the error is unknown and the PDF is usually created.
FIX or WORKAROUND
- Despite the error message, the PDF is created.
- Check the destination folder for presence of the PDF.
- Copy a known, working PDF to the destination folder.
- Rename the PDF file Printed.PDF.
- When saving the PDF, from Report Publisher select the Printed.PDF.
- Click Yes when prompted to overwrite.
- Open the Printed.PDF from the destination folder.
- Save As and name the PDF with the desired name in the desired location.
ADDITIONAL INFORMATION
Confirm if Terminal Server/Citrix have enough free space on the disk.
Confirm what is configured in System Environment Variables for TEMP folder.