No results or result set in Empower for PDA data - WKB9131
Article number: 9131
SYMPTOMS
- Sample set processes with no results
- Plenty of Tablespace in Project
- PDA Option is selected in the Project Properties
- Empower message center may show error "There are no channels specified to clear in method set"
ENVIRONMENT
- Empower 3 Personal
CAUSE
3D data without an Extracted Chromatogram in the Method Set
FIX or WORKAROUND
- Create a PDA type processing method.
- Derive a channel in the Method Set.
- Save the Method Set.
- Reprocess the Sample Set with the saved Method Set.