No results or result set in Empower for PDA data - WKB9131

Article number: 9131

SYMPTOMS

  • Sample set processes with no results
  • Plenty of Tablespace in Project
  • PDA Option is selected in the Project Properties
  • Empower message center  may show error "There are no channels specified to clear in method set"

ENVIRONMENT

  • Empower 3 Personal

CAUSE

3D data without an Extracted Chromatogram in the Method Set

FIX or WORKAROUND

  1. Create a PDA type processing method.
  2. Derive a channel in the Method Set.
  3. Save the Method Set.
  4. Reprocess the Sample Set with the saved Method Set.


PDAMS.jpg

ADDITIONAL INFORMATION

Empower PDA Software Getting Started Guide