Empower error messages "Raw data file XXXX does not match database entry" and "There is a time difference between the client and the server" - WKB90605

Article number: 90605

SYMPTOMS

  • Intermittent time difference when processing or acquiring data
  • Data set is not integrated when processing 
  • Errors "Raw data file XXXX does not match database entry" and "There is a time difference between the client and the server" in message center
  • Rebooting the server fixes the issue temporarily
  • DBRM seeing time stalls or reversals; scheduling suspended
    DBRM no longer seeing time stalls or reversals; scheduling resumed
    Warning: VKTM detected a time drift

ENVIRONMENT

  • Empower Enterprise 

CAUSE

The BIOS time zone does not match Oracle.

FIX or WORKAROUND

Correct the BIOS time zone to match Empower software.

ADDITIONAL INFORMATION


id90605, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP