Empower error messages "Raw data file XXXX does not match database entry" and "There is a time difference between the client and the server" - WKB90605

Article number: 90605

SYMPTOMS

  • Intermittent time difference when processing or acquiring data
  • Data set is not integrated when processing 
  • Errors "Raw data file XXXX does not match database entry" and "There is a time difference between the client and the server" in message center
  • Rebooting the server fixes the issue temporarily
  • DBRM seeing time stalls or reversals; scheduling suspended
    DBRM no longer seeing time stalls or reversals; scheduling resumed
    Warning: VKTM detected a time drift

ENVIRONMENT

  • Empower Enterprise 

CAUSE

The BIOS time zone does not match Oracle.

FIX or WORKAROUND

Correct the BIOS time zone to match Empower software.

ADDITIONAL INFORMATION