What does an Empower project backup integrity error mean? - "Error: Failed to read good raw datafile with ID" - WKB8913

Article number: 8913

ENVIRONMENT

  • Empower 2
  • Empower 3
  • NuGenesis 8 SDMS

ANSWER

This error means that the checksum stored in the database for the raw data files that was originally acquired does not match the checksum for the same raw data file during the project back up process. This means that the raw data files is missing or cannot be read.

This may be corrected if a good raw data file can be restored from a previous backup. These errors may be caused by corrupted, incomplete or aborted runs that did not complete to ther expected file size. Maybe loss of connection.

You can try to check the channel with the same ID number by taking it into review, also check if the run/injection was aborted. Compare the size of the channel in the project folder compared to the other similar channels.

ADDITIONAL INFORMATION

 