ICS Companion service does not start on LAC/E32 with Empower 3 installed - WKB88993

Article number: 88993

SYMPTOMS

  • ICS Companion service does not start
  • Waterservice, which depends on the ICS Companion, does not start
  • Deployment Manager "Repair" utility freezes

ENVIRONMENT

  • Empower 3 Enterprise
  • LAC/E32

CAUSE

Drivers from multiple driver pack versions are installed concurrently on the LAC/E32.

FIX or WORKAROUND

  1. Uninstall all drivers using the Deployment Manager.
  2. Reinstall the correct driver pack version.

ADDITIONAL INFORMATION

Try ICS Repair First