SYMPTOMS

  • All LAC/Es on the network system are running in buffer mode.
  • Cannot review recently acquired data.

ENVIRONMENT

  • Empower 3
  • LAC/E32

CAUSE

Data share is full. Navigate to data drive and look for available space.

FIX or WORKAROUND

  1. Increase space allowed to Data Share.
  2. Archive any remove information to create free space.

ADDITIONAL INFORMATION

 