No amounts are calculated in Empower - WKB87826

Article number: 87826

SYMPTOMS

  • No amounts are calculated in Empower

ENVIRONMENT

  • Empower 3

CAUSE

Peak Name, amounts or levels are not defined in the Amount Table for standards.

FIX or WORKAROUND

  1. Enter correct Peak Names, Amount values, and Levels in the SS amounts table
  2. Alternatively Enter correct Peak Names, Amount values and Levels, in the processing method Default Amounts Table.
  3. Reprocess.

ADDITIONAL INFORMATION

Note: Values in the Default Amounts tab of the processing method are superseded by any values in the sample set amounts table