Data is not pushing to the Empower server - WKB87703

Article number: 87703

SYMPTOMS

  • Data is not pushing to the Empower server
  • Data is buffering on the LAC/E32
  • Can ping the LAC/E32 IP address but not the LAC/E name
  • Cannot map to the file share from the LAC/E

ENVIRONMENT

  • Empower 3 Workgroup
  • Empower 3 Enterprise

CAUSE

The DNS server cannot resolve the LAC/E32 name.

FIX or WORKAROUND

Go to C > Windows > System32 > Drivers > ETC and open the Hosts file. Look to see if there is a line with LAC/E32 IP and name. If there is and there is a # sign in front of it, delete it. 