No amounts calculated for correctly identified peaks in Empower - WKB86336
Article number: 86336
SYMPTOMS
- No amounts calculated for correctly identified peaks in Empower
- Calibration standards were run prior to the unknowns, but in a different sample set
ENVIRONMENT
- Empower 3
CAUSE
The calibration standards were run in another project, so there is no calibration curve associated with the processing method being used to process the unknowns.
FIX or WORKAROUND
- Run the standards again in the correct project.
- Process the unknowns again.