Collected data not visible in Empower - WKB84764
Article number: 84764
SYMPTOMS
- Collected data is not visible in project window
- Run Samples shows sample set completed
- Run Samples displays chromatograms during acquisition
ENVIRONMENT
- Empower (all versions)
CAUSE
- The date on the PC is incorrect. Data has incorrect Acquisition Date.
- User is collecting the data in the parent project instead of the desired child project.
FIX or WORKAROUND
- Correct the date on the PC.
- Remove parent project access for users to prevent data generation in parent project.
ADDITIONAL INFORMATION
IT had recently worked on this computer and did not set the date correctly.