Agilent 7102A ELSD panel in Run Samples is dimmed (unavailable) - WKB84760
Article number: 84760
SYMPTOMS
- Agilent ELSD panel in Run Samples is dimmed (unavailable)
- No communication
ENVIRONMENT
- Empower 3
- LAC/E FR3
- ICF 3.1
- Agilent 1260
CAUSE
Communication with ELSD is not established.
FIX or WORKAROUND
- Remove the ELSD from the system configuration.
- Test Agilent 1260 without ELSD.
- If the 1260 system works, take the system offline.
- Configure the ELSD back into the system.
- Bring the system online.
ADDITIONAL INFORMATION
The order of configuration has made a difference in several systems with an Agilent 7102A ELSD.
