With Empower 3, SQT for Software shows only four pages of IQ report - WKB83523
Article number: 83523
SYMPTOMS
- Incomplete report - Running SQT for Software on a LAC/E shows only four pages of an IQ report; running Verify Files shows the complete report
ENVIRONMENT
- Empower 3
CAUSE
The NODE name in the Empower Configuration Nodes list is case-sensitive. You must make sure that it matches the LAC/E name exactly.
FIX or WORKAROUND
- As an Empower Administrator, from the server, delete the LAC/E from the Node List.
- Add the Node by browsing the network to find it.
- Rerun the SQT for Software IQ report.
ADDITIONAL INFORMATION