Agilent systems disconnect randomly during acquisition; process terminates due to unhandled exception error - WKB83434
Article number: 83434
SYMPTOMS
- Agilent systems disconnect randomly during acquisition
- Process terminates due to an unhandled exception error (exception info: System.OutOfMemoryException) message in Windows event application logs, followed by faulting module name: KERNELBASE.dll (faulting module path: C:\Windows\System32\KERNELBASE.dll)
- "Analysis aborted by another module" error message in the message center
- Antivirus is disabled
ENVIRONMENT
- Agilent Instrument Control Framework 3.0 and above
- Config 16 LAC/E
- Empower 3 Enterprise
CAUSE
Insufficient virtual memory
FIX or WORKAROUND
Increase the virtual memory as 4X the installed RAM.