Agilent 7890 reports an error during injection with Empower: Vial Attributes not valid - WKB80238

Article number: 80238

SYMPTOMS

  • 7890 reports an error during injection with Empower: Vial Attributes not valid

ENVIRONMENT

  • Empower 3

  • Agilent 7890 GC ICS V2.6

CAUSE

The injection volume in the samples table is set too high.

FIX or WORKAROUND

The maximum injection volume should not exceed 50% of the syringe volume.

If the injection volume exceeds 50%, install a larger syringe.

Define the correct syringe volume by clicking "Fr. Syringe" or "Bk. Syringe" in the Run Sample control panel.

ADDITIONAL INFORMATION

When using the 7890 without the Headspace, create a separate 7890 system in Empower.

See WKB62833 on Split injection technique.

See WKB62831 on Splitless injection technique.

See WKB62933 on PTV injection technique.

id80238, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP