Skip to main content
Waters

Every time when stop monitoring Agilent LC 1260 controlled by Empower, the message "Analysis aborted by another module" appears in the Message Center. - WKB80387

Article number: 80387

SYMPTOMS

  • Every time when stop monitoring Agilent LC 1260 controlled by Empower, the message "G1329B: DEAAC*****-Analysis aborted by another module" appeared in the Message Center.
  • Data acquires for few minutes and before completing the Run Time, moves to next Injection.
  • Sample set continues to next Injection 
  • Short run time on analysis aborted, but next injection starts without any delay

ENVIRONMENT

  • Empower 3
  • Agilent ICF Support Layer
  • .NET Framework 4.7

CAUSE

This is caused by the design of Agilent LC&CE Driver included in the bottom tier of Agilent ICF

FIX or WORKAROUND

CRI-2759 has been raised, not fixed with Agilent ICF Support Layer 3.5

There is no solution yet, but we can provide an Agilent instruction document about this issue.

See attached link

ADDITIONAL INFORMATION

id80387, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.