Empower 3 CE/CIA Time Corrected Area and Percent Time Corrected Area are not available as a Y Value in the Components tab of the processing method - WKB74649

Article number: 74649

SYMPTOMS

  • Time Corrected Area (Time Corr. Area) and % Time Corrected Area (% Time Corr. Area) are not available to select as a Y Value in the Components tab of the processing method
  • The CE/CIA option is turned on for the project, and this is seen when creating a CE or CIA processing method

ENVIRONMENT

  • Empower 3

CAUSE

This is a defect.

FIX or WORKAROUND

  1. Create a Peak custom field and set it equal to Time Corr. Area or % Time Corr. Area.
  2. Open the project after creating the new custom field and this field will be available for selection as a Y Value.

ADDITIONAL INFORMATION

PCS# 52189 was entered for this issue and is posted to the Waters website.

This issue has been addressed in Empower 3 SR5.