Empower 3 CE/CIA Time Corrected Area and Percent Time Corrected Area are not available as a Y Value in the Components tab of the processing method - WKB74649
Article number: 74649
SYMPTOMS
- Time Corrected Area (Time Corr. Area) and % Time Corrected Area (% Time Corr. Area) are not available to select as a Y Value in the Components tab of the processing method
- The CE/CIA option is turned on for the project, and this is seen when creating a CE or CIA processing method
ENVIRONMENT
- Empower 3
CAUSE
This is a defect.
FIX or WORKAROUND
- Create a Peak custom field and set it equal to Time Corr. Area or % Time Corr. Area.
- Open the project after creating the new custom field and this field will be available for selection as a Y Value.
ADDITIONAL INFORMATION
PCS# 52189 was entered for this issue and is posted to the Waters website.
This issue has been addressed in Empower 3 SR5.