Run Samples freezes when trying to set up an instrument method in Empower - WKB71895

Article number: 71895

SYMPTOMS

  • Run Samples freezes when trying to set up an instrument method
  • Preconfiguration matches the Agilent 1260 physical configuration
  • Method resolution errors appear in the message center
  • A new instrument method sets up successfully in Run Samples

ENVIRONMENT

  • Empower 3 
  • Agilent 1260 Infinity LC
  • Agilent Instrument Control Framework
  • 2695 Separations Module

CAUSE

The instrument method in use was created with a different detector

The instrument method must match the instrument configuration.

FIX or WORKAROUND

Create a new instrument method with the correct components.

ADDITIONAL INFORMATION

 