Boot -9 error in the node properties under BusLAC/E Status in Breeze - WKB71544

Article number: 71544

SYMPTOMS

  • Boot -9 error in the Node Properties under BusLAC/E Status

ENVIRONMENT

  • Empower 
  • Breeze
  • BusLAC/E card

CAUSE

Dirty contacts

FIX or WORKAROUND

  1. Power-off the PC.
  2. Remove the BusLAC/E card.
  3. Clean the BusLAC/E card contacts with an eraser.
  4. Insert the BusLAC/E card.
  5. Power-on PC.