Instrument server crash with ntdll.dll error during an Empower sample set - WKB71513
Article number: 71513
SYMPTOMS
- Instrument server crashes with an ntdll.dll error during an Empower sample set
- Increasing virtual memory does not resolve the issue
ENVIRONMENT
- Empower 3 Feature Release 2
- ICF 3.0
- Agilent 1290 Infinity II
CAUSE
This is a defect.
FIX or WORKAROUND
The defect CRI-431 has been fixed in Agilent Instrument Control Framework Release version 3.2
Workaround:
1. Have no condition column or equilibrate function in the sample set.
2. If condition column or equilibrate function must be used, create a new sample set with an inject sample (Blank) as the first line, followed by a condition or equilibrate function column. Include no more than one condition column in a sample set.
ADDITIONAL INFORMATION
Additional information regarding the instrument server crash with Empower 3 FR2 and Agilent Infinity II systems can be found here.