How to configure plates for an Alliance 2795 - WKB69946

Article number: 69946

OBJECTIVE or GOAL

Configure plates for an Alliance 2795.

ENVIRONMENT

  • Alliance 2795
  • Empower 3

PROCEDURE

  1. Open the Run Samples window in Empower. This may also be done from the Quickstart interface.
  2. At the top of the window, click Edit > Plates.
  3. In the new window, select "2790 Layout".
  4. Click the number corresponding to the plate position that will be used. There are four available positions.
  5. The drop-down list for that position becomes active.
  6. From the drop-down list, select the plate type that will be used.
  7. Click OK at the bottom of the window.

ADDITIONAL INFORMATION

 