Duplicate results generated for all results in Empower - WKB6904
Article number: 6904
SYMPTOMS
- When performing batch processing, duplicate results generated for all data in same sample set
- All users reporting same issue in same sample set
- Other sample sets within same project working as intended
- Label reference column is identified correctly
- Sample Type column is identified correctly
ENVIRONMENT
- Empower 3
CAUSE
The Processing column in the sample set method is incorrectly set up for Quantitate.
FIX or WORKAROUND
- Go to Sample Set tab, right-click and go to Alter Sample.
- Type "Do not process or report" for all samples.
- Add a Calibrate line at the end, and use S* for Label reference.
- Set processing column to Normal.
- Add a function called Quantitate next and use C*, U* for Label reference.
- Set processing column to Normal.
- At the last line function, type "summarize custom fields" if you have custom fields.
- Set processing column to Normal.
- Save sample set and reprocess.
ADDITIONAL INFORMATION
