Export Spawn Process error occurs during project backup in Empower - WKB6646
Article number: 6646
SYMPTOMS
- "Export Spawn Process failed"
- No projects can be backed up
ENVIRONMENT
- Empower 3
CAUSE
- The PC where the backup is being performed has a full C drive.
- The server has not been restarted for more than a month. The system is unable to allocate session/process/memory required to perform the backup operations at database level.
FIX or WORKAROUND
- Clear memory on the PC, and then back up projects.
- If the issue stems from the server not being restarted, restart the database server to clear memory.
Note: This behavior may not be observed with cold backup enabled servers because during cold backup all sessions/process and memory are cleared. Because RMAN backup is being used, until the database is restarted, sessions will not be cleared. Servers should be restarted at least once a month.