Error: "You do not have the necessary options enabled to display some or all of the data" when reviewing QDa data - WKB6636
Article number: 6636
SYMPTOMS
- Error: "You do not have the necessary options enabled to display some or all of the data"
- Cannot view QDa data
ENVIRONMENT
- Empower 3
CAUSE
Mass spectrometry option is not enabled in the project.
FIX or WORKAROUND
- Open Configure the system.
- Right-click the project and choose Properties.
- For enabled options, select Mass Spectrometry.
ADDITIONAL INFORMATION