How to Collect the Agilent ICF logs and Lab Advisor diagnostic buffer Logs - WKB65971
OBJECTIVE or GOAL
Capture all of the available diagnostic data and logs immediately after you encounter an instrument error.
ENVIRONMENT
- Empower
- ICFSL v3.0 and above
PROCEDURE
The Agilent diagnostic buffers and logs can be captured in multiple ways.
NOTE: Collect logs before restarting the LAC/E or acquisition client to help investigate the root cause.
IMPORTANT - Collect the Agilent Diagnostic logs immediately after the incident and before rebooting the instrument modules.
For ICFSL v3.0 - v3.1
- Run the AgilentLogfileCollecter.exe - C:\Empower\Instruments\AgilentLC\Support\Agilent.DriverLogFileCollector.exe
- Use Agilent LabAdvisor to collect the Agilent Diagnostic logs
Additionally, in the Run Samples window there is a tab for diagnostic logs; print or copy to Notepad and save.
For ICFSL v3.2 and above see How to enable or disable Agilent third-party logs - WKB227265 - Waters
ADDITIONAL INFORMATION
IMPORTANT - The diagnostic logs are immediately overwritten when the instrument modules are rebooted. If the logs are not captured, they will be lost, greatly reducing the chances of identifying the root cause of the instrument error. The files will be zipped up and may go to the documents folder in Windows.