Skip to main content
Waters

Altering an ICF-based Agilent LC instrument method in Empower results in an instrument failure error message - WKB65729

Article number: 65729

SYMPTOMS

  • A newly created ICF based Agilent LC instrument method works properly.
  • Editing the newly created ICF based Agilent LC instrument method results in an instrument failure message when using the method.
  • Auxiliary Channels tab is blank.

ENVIRONMENT

  • Empower 3
  • ICF Support v2.2
  • ICF Support v3.0

CAUSE

This behavior is documented as CRI-330. The problem appears when an Agilent Universal Interface Box (UIB) is present in the system configuration.

FIX or WORKAROUND

Remove the Universal Interface Box (UIB) from the chromatographic system configuration.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.