Memory Buffer error from e-SAT/IN with ICS 1.30 and Empower - WKB65702

Article number: 65702

SYMPTOMS

  • Memory Buffer error in Empower message center. 
  • The socket failed while waiting for incoming data with the following error code: WinSock Error #0.
  • Run stops after first six minutes.

ENVIRONMENT

  • e-SAT/IN Module Instrument Control Software Version 1.30
  • Firmware: 3.2
  • Empower 2 Feature Release 5 Enterprise

CAUSE

Defect PCS 49400

Abstract / Summary

The version 1.30 Instrument Control Software (ICS) for the e-SAT/IN module suffers from a vulnerability that may produce a false positive “Memory Buffer Error” message when running at low data rates. The probability of receiving a false positive error message is highest at a sampling rate of two points per second. Version 2.6 or 3.2 e-SAT/IN firmware cannot produce a real memory buffer error message and any such messages generated while using e-SAT/IN firmware version 2.6 or 3.2 should be considered false positive messages. You should investigate any memory buffer error found while using e-SAT/IN firmware version 2.5 or 3.1.

FIX or WORKAROUND

  1. If possible, increase the sampling rate (the number of points per second) being used to collect data. Alternately, you can operate e-SAT/IN modules using version 2.6 or 3.2 firmware with ICS version 1.20 without the risk of a real memory buffer error.

ADDITIONAL INFORMATION

 