Plates missing from the "assign well plates" drop-down menu for Agilent G7167B Multisampler in Empower - WKB64861
Article number: 64861
SYMPTOMS
- Plates missing from the "assign well plates" drop-down menu in Run Samples when trying to select a plate
- Importing the plates from the driver does not allow the plate drop-down menu to populate with the plate configurations
- Cannot run a sample set because the plates need to be configured with the Multisampler
- Deleting and recreating the system via the Agilent PreConfiguration utility does not resolve the issue
ENVIRONMENT
- Empower
- Agilent LC 1290 Infinity II
- Agilent LC G7167B Multisampler
- ICF 3.0
CAUSE
The drawers were configured incorrectly because the sample plates were not removed.
FIX or WORKAROUND
- Configure the Multisampler as described in WKB24672.
- Follow the instructions and prerequisites outlined in the drawer configuration tool.