Incomplete run over weekend with Empower and Agilent LC 1100 - WKB64809
Article number: 64809
SYMPTOMS
- Sequence is incomplete; random drop after 38 injections
- Recover.log is clean
- Message center is clean
- Windows event logs are clean
ENVIRONMENT
- Empower 3 Feature Release 1 Enterprise
- Module G1311A; firmware version A.05.11
- Module G1313A; firmware version A.05.11; syringe size 100
- Module G1314A; firmware version A.05.11
- Module G1316A; firmware version A.05.11
- Agilent 1100
- Agilent LC Instrument Control Software Version 1.06
CAUSE
- On the LAC/E, two systems are connected.
- System 1 shows the communication failure message "IOThreadProc: Agilent LC Communication failure waiting for single object.A1100_TIMEOUT", which affected System 2 running on the LAC/E.
- This timeout error can happen while running, when any parameter is unable to reach the set point in a defined time.
FIX or WORKAROUND
- Check conditions prior to run.
- Don't connect multiple systems at the same time.
ADDITIONAL INFORMATION
Cause is determined from the Empower Acquisition log (Open injection via the report publisher > Look for the Acquisition log in report on the left-hand side).
id64809, comms