SYMPTOMS

  • Peaks identified incorrectly
  • Previous peaks that don't have a close retention time are labeled

ENVIRONMENT

  • Breeze 2
  • Empower 3

CAUSE

Incorrect retention time in the processing method

FIX or WORKAROUND

  1. Update the processing method with correct retention time for peak/compound of interest.
  2. Reprocess.
  3. Verify correct peak identification.

ADDITIONAL INFORMATION

 

id64419, BREEZESW, SUPBZ

Not able to find a solution? Click here to request help.