Incorrect peak identification with Breeze 2 - WKB64419
Article number: 64419
SYMPTOMS
- Peaks identified incorrectly
- Previous peaks that don't have a close retention time are labeled
ENVIRONMENT
- Breeze 2
- Empower 3
CAUSE
Incorrect retention time in the processing method
FIX or WORKAROUND
- Update the processing method with correct retention time for peak/compound of interest.
- Reprocess.
- Verify correct peak identification.
ADDITIONAL INFORMATION