Empower method is unlocked after selecting Cancel in the dialog for typing comments - WKB57447
Article number: 57447
SYMPTOMS
- System policies are set for the Empower project to require comments for method changes
- Select Method Properties for a locked method, click the Unlock button
- Cancel from the Comments dialog, where comments and password for identity confirmation may or may not have been specified
- Select either Cancel or OK from the method properties
- The method is unlocked
- There is an Unlocked Method audit record, but the method version and comments are the same as for the last saved method
- There is no Modified Method audit record, which is normally expected after the unlock action when the method is saved as a new version
ENVIRONMENT
- Empower 3 SR3 and above
CAUSE
This defect was filed as CRI-218.
FIX or WORKAROUND
The method must be locked again if that is the necessary setting.
ADDITIONAL INFORMATION
This defect is fixed in Empower 3.8.0.