Sample set does not load in the Run Samples window of Empower - WKB5720
Article number: 5720
SYMPTOMS
- In the Run Samples window, the load icon is clicked and Load using a previously created sample set method is selected. After OK is clicked, no sample sets populate the window
ENVIRONMENT
- Empower 3
CAUSE
- The project's maximum number of preferences have been saved to the project.
FIX or WORKAROUND
Replace preferences with those of a working project:
- Go to Configuration Manager and click on View > Copy Preferences.
- Copy the preferences of a working project into the project that has the problem.
ADDITIONAL INFORMATION
When copying the preferences, you must select a user source and a user, in addition to the project preferences. (The user source and the user can be the same if you do not want to change the user preferences.)