"ICS Repair Status : Please wait while completing configuration" message at login to Empower - WKB54395
Article number: 54395
SYMPTOMS
- "ICS Repair Status : Please wait while completing configuration" message at login
- Uninstallation and reinstallation of drivers temporarily resolves issue, but it returns
ENVIRONMENT
- Empower 3, all platforms
CAUSE
Antivirus exclusions were not done at time of installation.
FIX or WORKAROUND
- Use the Repair function in the ICS component by launching it as a full admin.
- Make exceptions to antivirus scanning (see below).
Here are the recommended folders to exclude from real-time scanning (include subfolders if given the option):
- C:\Empower
- C:\Program Files (x86)\Waters Instruments
- C:\Program Files (x86)\Oracle
- C:\Program Files\waters instruments