Changes made to sample set method are not saved using "Save Preferences" in Empower - WKB52662
Article number: 52662
SYMPTOMS
- Changes made to sample set method are not saved using "Save Preferences" in Empower
- Table is modified to display in a specific order and "Save Preferences" is applied; when the run window is closed and reopened, the table reverts back to the original view
ENVIRONMENT
- Empower 3 FR3
CAUSE
Changes made to the sample set method tables are saved with "Save Table Preferences", not "Save Preferences".
FIX or WORKAROUND
Use "Save Table Preferences" to save the changes.
ADDITIONAL INFORMATION
This action requires the "Save Named Table Preferences" privilege found in User Type Properties on the Management tab.