Altered Sample audit trail is not associated with the data that was altered using "View As" from the data - WKB52802
Article number: 52802
SYMPTOMS
- From the project window, viewed a Sample Set as Injections
- Selected one injection, made a change using the Alter Sample tool, and saved the change
- Selected the audit trails tab and the "Altered Sample" audit trail was seen, as expected
- When performing a "View As Audit Records" from the sample set, the injection, or the channel associated with the altered sample, the audit record is not shown
ENVIRONMENT
- Empower 3
CAUSE
This is a defect.
FIX or WORKAROUND
Selecting the sample set, instead of a single injection, and then altering one injection results in an "Altered Sample Set" audit record that will be associated with the sample set.
ADDITIONAL INFORMATION
PCS# 56628 ("Alter sample for single injection doesn't link audit trail to any data") has been entered.