Unknown channels are collecting using Agilent Intuvo 9000 GC with Empower 3 - WKB51317
Article number: 51317
SYMPTOMS
- Running a sample set with more than one injection per line times out at the start of the second injection
- Multiple unknown channels are collecting for the second injection of the line in the sample set
- Stopping the sample set and restarting it at the second injection of the line brings the same result: the injection runs as normal, but then when going to the next injection of the line in the sample set, it times out, resulting in unknown channels
ENVIRONMENT
- Empower 3 Feature Release 4
- Agilent Instrument Control Framework (ICF) 3.0
- Agilent Instrument Control Framework (ICF) Update A02.05
- Agilent Intuvo 9000 GC
CAUSE
Defect (PCS #56603)
FIX or WORKAROUND
Include only one injection per line of the sample set.
Upgrade to ICF Support Layer v3.2 and Agilent ICF 2.6 Update 2.
ADDITIONAL INFORMATION
id51317, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP