Unknown channels are collecting using Agilent Intuvo 9000 GC with Empower 3 - WKB51317
Article number: 51317
SYMPTOMS
- Running a sample set with more than one injection per line times out at the start of the second injection
- Multiple unknown channels are collecting for the second injection of the line in the sample set
- Stopping the sample set and restarting it at the second injection of the line brings the same result: the injection runs as normal, but then when going to the next injection of the line in the sample set, it times out, resulting in unknown channels
ENVIRONMENT
- Empower 3 Feature Release 4
- Agilent Instrument Control Framework (ICF) 3.0
- Agilent Instrument Control Framework (ICF) Update A02.05
- Agilent Intuvo 9000 GC
CAUSE
Defect (PCS #56603)
FIX or WORKAROUND
Include only one injection per line of the sample set.
Upgrade to ICF Support Layer v3.2 and Agilent ICF 2.6 Update 2.
ADDITIONAL INFORMATION