Skip to main content
Waters

Unknown channels are collecting using Agilent Intuvo 9000 GC with Empower 3 - WKB51317

Article number: 51317

SYMPTOMS

  • Running a sample set with more than one injection per line times out at the start of the second injection
  • Multiple unknown channels are collecting for the second injection of the line in the sample set
  • Stopping the sample set and restarting it at the second injection of the line brings the same result: the injection runs as normal, but then when going to the next injection of the line in the sample set, it times out, resulting in unknown channels

ENVIRONMENT

  • Empower 3 Feature Release 4
  • Agilent Instrument Control Framework (ICF) 3.0
  • Agilent Instrument Control Framework (ICF) Update A02.05
  • Agilent Intuvo 9000 GC

CAUSE

Defect (PCS #56603)

FIX or WORKAROUND

Include only one injection per line of the sample set.

Upgrade to ICF Support Layer v3.2 and Agilent ICF 2.6 Update 2.

ADDITIONAL INFORMATION

 

id51317, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.