Unable to export results in Empower - WKB4840
Article number: 4840
SYMPTOMS
- No errors observed.
- Data appears normal in review.
- Results are generated but do not export into the designated folder.
ENVIRONMENT
- Empower 3 Personal.
CAUSE
Project Tablespace is too low.
FIX or WORKAROUND
- Click Configure the System to open Configuration Manager.
- Highlight the Project name.
- Right-click and select Properties.
- Increase Quota.
ADDITIONAL INFORMATION