Article number: 46878
SYMPTOMS
- The all peaks table group contains information that does not match the data reported when reviewing results
ENVIRONMENT
- Empower 3
CAUSE
The retention time shifted and the peak of interest is not integrated.
FIX or WORKAROUND
- Correct the retention time in the processing method to within an acceptable range.
- Reprocess the sample set.
ADDITIONAL INFORMATION
A large shift in retention time is more likely a hardware issue.