Is it possible to enable a headspace 7697A to inject with a vial start and end range in the instrument method different than the value in the sample set? - WKB46821
Article number: 46821
ENVIRONMENT
- Empower 3 Feature Release 2
- Agilent 7697A headspace
- Agilent 7890 GC
- Headspace Control Option 3.0
ANSWER
No, it is not possible. The Start Vial and End Vial positions for the 7697A need to match what is in the Sample Set method and Empower Instrument method.
ADDITIONAL INFORMATION
This is addressed in Instrument Control Framework 3.0 as vial numbers are no longer required in the instrument method. When running samples, Empower software derives the vial range from the sample set method.
For additional information, see the release notes for Instrument Control Framework 3.0.