Alliance Flow Rate Change button grayed out and cannot change the flow rate - WKB46031
Article number: 46031
SYMPTOMS
- On the analysis window of Empower, the Alliance Flow Rate Change button is always grayed out and cannot change the flow rate.
ENVIRONMENT
- Empower
- Alliance
CAUSE
The privilege of "Allow Interactive Sys Changes" for user's data acquisition privilege is turned off.
FIX or WORKAROUND
Turn on the privilege of "Allow Interactive Sys Changes" for the user.