Is it possible to disable the Stop time and Post time in the Instrument Method when running Agilent LC with Empower? - WKB4539

Article number: 4539

ENVIRONMENT

  • Empower 3.
  • Agilent Instrument Control Framework Support Version 2.2.
  • Agilent 1100 and 1200 series.

ANSWER

  • No, you cannot. PCS 55397 is logged in to "disable stop time in Agilent instrument method".
  • In Empower, the Sample Set Method overrides whatever volume is incorporated into the Instrument Method.
 You can also refer to KCS article 3337 regarding disable injection volume in Agilent Instrument Method.  

ADDITIONAL INFORMATION

id4539, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP