Skip to main content
Waters

Does Empower use the injection volume from Instrument Method or Sample Set Method in Agilent LC? - WKB3337

Article number: 3337

ENVIRONMENT

  • Empower 3 all platforms
  • All Agilent LC's

ANSWER

In Empower, the Sample Set Method overrides whatever volume was incorporated into the Instrument Method.

ADDITIONAL INFORMATION

In Empower, the injection volume field setting is set via the sample set method. However, when Agilent ICF is installed, in the instrument method, there is now an option for the injection volume to be set here, too. Even though there is now an additional place to set the injection volume with the Agilent ICF, the injection volume function in Empower remains the same. It is controlled by the sample set method. Therefore, the sample set method overrides whatever volume was incorporated into the instrument method.

 

Injection volume.PNG

EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.