Intermittent communications issues with instruments - WKB45401
Article number: 45401
SYMPTOMS
- The QDa had intermittent communication issues with Empower during overnight runs
- The QDa had to be restarted to resume communications
- QDa communication issues appeared to stop, but then the ACQUITY Column Manager had similar issues
- Intermittent losses of communication with LCMS system
ENVIRONMENT
- Empower 3
- UNIFI 1.9.3 Workstation
- Xevo G2-XS
- ACQUITY UPLC I-Class
CAUSE
The computer had the power-saving mode set to turn off the hard drive after 30 minutes of inactivity.
FIX or WORKAROUND
In Windows Power Options, change the hard disk power-saving mode to never turn off the drive.
ADDITIONAL INFORMATION
Waters-built PCs are set, by default, to the Waters power plan under Control Panel\All Control Panel Items\Power Options. Customer-provided PCs may instead be using other power plans (e.g., Balanced) in which the advanced power settings are configured incorrectly.