How to install and configure the Buslace driver in Empower - WKB4389
OBJECTIVE or GOAL
To install a busLAC/E driver in Empower, based on the operating system.
ENVIRONMENT
- Empower 2
- Empower 3
- Breeze
- Windows XP
- Windows 7
- Windows 10
PROCEDURE
To install and configure the busLAC/E driver on Windows XP:
- Right-click My Computer, and then select Manage.
- On the Computer Management page, click Computer Management (Local) > System Tools > Device Manager. In the right-hand pane, locate Other devices > PCI Device.
- Right-click PCI Device, and then select Properties.
- In the General tab on the PCI Device Properties page, click the Reinstall Driver button. The Found New Hardware Wizard page appears.
- On the Found New Hardware Wizard page, select “Yes, this time only”, and then click Next.
- Select “Install from a list or specific location (Advanced)”, and then click Next.
- Select “Search for the best driver in these locations” and select the checkbox “Include this location in the search”.
- Click Browse and include the following directory in the search: C:\Empower\BuslaceDrivers.
- Click Next.
Result: The BusLAC/E Driver installs. It appears as “Buslace PCI KMDF Device Driver” under BusLAC/E Drivers in Device Manager.
To install and configure the busLAC/E driver on Windows Vista:
- Right-click Computer, and then select Manage.
- On the Computer Management page, click Computer Management (Local) > System Tools > Device Manager. In the right-hand pane, locate Other devices > PCI Device.
- Right-click PCI Device, and then select Properties.
- In the General tab on the PCI Device Properties page, click Reinstall Driver. The Found New Hardware - PCI Device page appears.
- On the Found New Hardware - PCI Device page, select "Browse my computer for driver software (advanced)."
- Under “Browse for driver software on your computer”, click Browse and select the directory C:\Empower\BuslaceDrivers.
- Select the checkbox “Include subfolders”, and then click Next.
- When the Windows Security message “Would you like to install this device software?” appears, click Install.
Result: The BusLAC/E Driver installs. It appears as “Buslace PCI KMDF Device Driver” under BusLAC/E Drivers in Device Manager.
To install and configure the busLAC/E driver on Windows 7:
- Right-click Computer, and then select Manage.
- On the Computer Management page, click Device Manage (under Computer Management (local) > System Tools).
- In the right-hand pane, right click Other Devices > PCI Device, and then select Update Driver Software.
- On the Update Driver Software - PCI Device page ("How do you want to search for driver software?"), select "Browse my computer for driver software" to manually install the busLAC/E drivers.
- On the Update Driver Software - PCI Devices page ("Browse for driver software on your computer"), click Browse.
- In the Browse for Folder dialog box, browse to X:\Empower\BuslaceDrivers64, where X:\ is the drive where you installed Empower, and then click OK.
- On the Update Driver Software - PCI Device page ("Browse for driver software on your computer"), ensure that the path is correct, and then click Next to start the driver installation.
- When the "Windows has successfully updated your driver software" screen appears, click Close.
Result: The Device Manager now shows the BusLACE PCI card listed under Waters Instrument Control Devices.
To install and configure the busLAC/E driver on Windows 10:
- In Windows 10, position the pointer in the bottom-left corner of the screen, rightclick
- Start, and click Computer Management.
- On the Computer Management page, click Device Manager (under Computer Management (Local) > System Tools).
- In the right-hand pane, right-click Other Devices > PCI Device, and then select Update Driver Software.
- On the Update Driver Software - PCI Device page (“How do you want to search for driver software?”), select Browse my computer for driver software, to manually install the busLAC/E drivers.
- On the Update Driver Software - PCI Device page (“Browse for driver software on your computer”), click Browse.
- In the Browse for Folder dialog box, browse to X:\Empower\BuslaceDrivers64, where X:\ is the drive where you installed Empower, and click OK.
- On the Update Driver Software - PCI Device page (“Browse for driver software on your computer”), ensure that the path is correct, and then click Next.
Result: The driver installation starts.
- When the Windows has successfully updated your driver software screen appears, click Close.
Result: The Device Manager now shows the BusLAC/E PCI card listed under Waters Instrument Control Devices.
ADDITIONAL INFORMATION
For Breeze 2, Please see the Installation and Configuration guide, page 3-5 "Installing and configuring a busLAC/E driver"