No results are processed in Empower 3 - WKB4031
Article number: 4031
SYMPTOMS
- Processing a sample set yields no result set.
ENVIRONMENT
- Empower 3
CAUSE
Sample Set linkage is corrupted.
FIX or WORKAROUND
- Verify that the free project tablespace is at least 10 Mb. If not, increase it by another 50 Mb in Configure the System, Project Properties.
- If you are still unable to process the Sample Set as a whole, process the Sample Set's channels individually.
- Acquire the next Sample Set using a new Sample Set Method from scratch instead of using the old Sample Set Method.