How to derive a channel in Empower. - WKB3852
Article number: 3852
OBJECTIVE or GOAL
This procedure is used to extract a two-dimensional chromatogram from a three-dimensional acquisition. This is useful when only 3D data was originally acquired by the PDA detector.
ENVIRONMENT
- Empower
- Empower 2
- Empower 3
PROCEDURE
- Open the Method Set used to acquire the data.
- Ensure that the Instrument Method listed at the top is a PDA method.
- Go to the top and choose "File - Create New Derived Channel".
- In the window that opens, click the Channel dropdown.
- Select "PDA: Single Wavelength".
- Enter the wavelength you wish to view.
- Click "OK" at the bottom of the window.
- Specify a name for this channel.
- Click "OK" at the bottom of the window.
- You will now see the channel listed in the table at the center of the Method Set Editor window.
ADDITIONAL INFORMATION
You may choose a different option in step 5 above. For instance, you could select to extract an MS TIC rather than a PDA single wavelength, if you had collected MS data to begin with.